What are the maintenance techniques for salt spray testers?

What are the maintenance techniques for salt spray testers?

First, the amount of spray is insufficient

cause of issue:

1. The spray regulator is placed too low;

2. The glass filter in the preheating tank is blocked;

3. The pressure setting is too low.

Approach:

1. Raise the spray regulator;

2. Clean the glass filter;

3. Adjust the pressure regulating valve to a pressure of 1kg per square centimeter. The air compressor is marked with a pressure regulating valve and adjusted to a pressure of 2kg per square centimeter.

Second, when the water level is low, the warning light is on.

Cause: The water level is too low.

Treatment: Check if the test chamber or saturated tank is adequately water level.

Third, there is a normal spray and the air compressor is not running.

The cause of the malfunction: The air compressor itself has a self-protection function.

Treatment: Use as usual without affecting the test process.

Fourth, can not spray ps: nozzle air pipe and suction pipe device method (straight pipe is the suction pipe; l type is the air pipe)

cause of issue:

1. The air compressor is not running;

2. The main switch of the air compressor outlet is not open;

3. Solenoid valve failure;

4. The pressure gauge is faulty or the pressure is too low;

5. Electromagnetic contactor failure;

6. The nozzle is blocked.

Approach:

1. Turn on the air compressor button;

2. Turn on the air compressor main switch;

3.4.5. Notify the repair company to repair the door;

6. Remove the nozzle and clean it (please disassemble it carefully).

Five, can not run after turning on the power

Cause of failure: When the water level in the heating tank is too low, the power supply will be cut off.

Treatment method: Add the water level of the heating tank to the normal condition.

Six, when the temperature controller displays hhh, lll

The cause of the fault: the temperature controller is faulty and the temperature sensing rod is faulty.

Treatment method: notify the maintenance company to repair the door.