Because the salt spray test machine has been doing salt spray aging test for a long time, the bad environment is easy to cause large and small failures of the salt spray test machine. And processing methods.
1. The laboratory cannot rise to the set temperature
cause of issue:
1. The temperature of the laboratory temperature controller is set too low;
2. The laboratory safety protection switch is set too low;
3. Heating system failure;
4. Electromagnetic relay failure;
5. The controller is faulty.
Approach:
1. Set the temperature controller to the desired temperature;
2. Set the safety protection switch to the required temperature;
3.4.5. Notify the maintenance company for home repair.
2. The temperature of the saturation barrel cannot rise to the set temperature
cause of issue:
1. The temperature of the saturation barrel temperature controller is too low;
2. The saturation barrel safety protection switch is set too low;
3. Heating system failure;
4. Electromagnetic relay failure;
5. The controller is faulty.
Approach:
1. Set the temperature controller to the desired temperature;
2. Set the safety protection switch to the required temperature;
3.4.5. Notify the maintenance company for home repair.
3. Insufficient spray volume
cause of issue:
1. The spray regulator is placed too low;
2. The glass filter in the preheating tank is blocked;
3. The pressure setting is too low.
Approach:
1. Turn up the spray regulator;
2. Clean the glass filter;
3. Adjust the pressure regulating valve to a pressure of 1Kg per square centimeter. The air compressor is marked with a pressure regulating valve to adjust the pressure to 2Kg per square centimeter.
4. When the low water level alarm light is on
Cause: The water level is too low.
Treatment method: check whether the water level in the test room or saturation tank is sufficient.
5. There is a normal spray and the air compressor is not running
Cause of failure: The air compressor itself has a self-protection function.
Treatment method: Use as usual, without affecting the test process.
Six, unable to spray
cause of issue:
1. The air compressor is not running;
2. The main switch of the air compressor outlet is not turned on;
3. Solenoid valve failure;
4. The pressure gauge is faulty or the pressure is too low;
5. Electromagnetic contactor failure;
7. The nozzle is blocked.
Approach:
1. Turn on the air compressor button;
2. Turn on the air compressor main switch;
3.4.5. Notify the maintenance company to come home for repair;
6. Remove the nozzle for cleaning (please disassemble carefully).
8. Cannot run after turning on the power
Cause of failure: When the water level in the heating tank is too low, the power supply for operation will be cut off.
Treatment method: increase the water level of the heating tank to normal conditions.
Nine, when the temperature controller displays HHH, LLL
Causes of failure: temperature controller failure, temperature sensor failure.
Solution: notify the maintenance company to repair